Commissioning
The device automatically locks the write-protected parameters again if a key is not pressed for
10 minutes in the navigation and editing view. The device locks the write-protected
parameters automatically after 60 s if the user skips back to the operational display mode
from the navigation and editing view.
• If write access is activated via access code, it can be also be deactivated only via the
access code (→  32).
• The user role with which the user is currently logged on via the local display (→  32)
is indicated by the Access status display parameter. Navigation path: Display/
operation → Access status display
9.6.2
Write protection via write protection switch
Unlike write protection via user-specific access code, this allows write access to the entire
operating menu - other than the Contrast display parameter - to be locked.
The parameter values are now read only and cannot be edited any more (exception Contrast
display):
• Via local display
• Via service interface (CDI)
• Via HART protocol
1.
Loosen the securing clamp.
2.
Unscrew the electronics compartment cover.
3.
Pull out the display module with a gentle rotational movement. To make it easier to
access the lock switch, attach the display module to the edge of the electronics
compartment.
38
CDI
WP
SIM
XXXXXXXXXXXXX
Proline Prosonic Flow B 200
OFF
ON
3 mm
A0013768
Endress+Hauser