1.888.610.7664
non-compensated or poorly compensated probe may cause measurement errors. The
compensation procedure is as follows:
1. Perform Step 1, 2, 3 and 4 in Function Inspection.
2. Check the displayed waveforms and compare them with the waveforms shown in
Figure 5.6
.
Over compensated
3. Use the probe compensation adjustment tool provided in the accessories to adjust
the low-frequency compensation adjustment hole on the probe until the displayed
waveform is consistent with the "Perfectly compensated" waveform shown in the
above figure.
www.calcert.com
15
10
20
5
25
0
30
Perfectly compensated
Figure 5.6 Probe Compensation
To Prepare for Use
Under compensated
sales@calcert.com