Calibration Data - Xylem YSI pH 1310P Operation Manual

Ph meter
Table of Contents

Advertisement

pH value
Displaying the cali-
bration data
Displaying the cali-
bration data storage
Calibration evalua-
tion
32

5.2.6 Calibration data

The calibration record is automatically transmitted to the interface
after calibrating.
The calibration record of the last calibration is to be found under the menu item,
<F1> / [Menu]Calibration / Calibration record. To open it, press the <CAL__>
key in the measured value display.
Subsequently, you can transmit the displayed calibration data to the interface,
e. g. to a PC, with <F2>[USB output].
The calibration records of the last calibrations (up to 10) are available in the
menu, <F1>/[Menu] / Calibration / Calibration data storage and in the menu,
<F1__>/[Menu] / Storage & config/Data storage / Calibration data storage.
Menu item
Calibration /
Calibration data stor-
age /Display
or
Data storage /
Calibration data stor-
age /Display
Calibration /
Calibration data stor-
age /
Output to USB
or
Data storage /
Calibration data stor-
age /Output to USB
After calibrating, the meter automatically evaluates the calibration. The zero
point and slope are evaluated separately. The worse evaluation of both is taken
into account. The evaluation appears on the display and in the calibration
record.
Setting/func-
Description
tion
-
Displays the calibration record.
Further options:
 Scroll through the calibration
records with <><>.
 Output the displayed calibra-
tion record to the interface with
<F2>/[USB output].
 Quit the display with <F1>/
[Back] or <MENU/ENTER>.
 Switch directly to the mea-
sured value display with <M>.
-
Outputs the calibration records to
the interface.
TruLab pH 1310
ba76156e01
03/2014