pH value
30
22.
Set the nominal buffer value for the measured temperature with
<
><
>.
23.
Accept the calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage display).
24.
If necessary, use <M> to finish the calibration.
The calibration record is displayed.
or
Continue calibrating using the next buffer with <ENTER>.
After the fifth buffer has been measured the calibration is automat-
ically finished. Then the calibration record is displayed.
The calibration line is determined by linear regression.
5.2.5 Calibration points
Calibration can be performed using one to five buffer solutions in any order
(single-point to five-point calibration). The meter determines the following
values and calculates the calibration line as follows:
Calibration
1-point
2-point
3-point to 5-
point
You can display the slope in the units, mV/pH or % (see section
7.1.1 S
ETTINGS FOR P
5.2.6 Calibration data
The calibration record is automatically transmitted to the interface
after calibrating.
Determined
Displayed calibration data
values
Zero point = Asy
Asy
Slope = Nernst slope
Zero point = Asy
Asy
Slp.
Slope = Slp.
Zero point = Asy
Asy
Slp.
Slope = Slp.
The calibration line is calculated by
linear regression.
H
MEASUREMENTS
HandyLab 600
(-59.2 mV/pH at 25 °C)
, page 40).
ba77064e04
08/2021