Calibration Points; Calibration Data - Xylem YSI TruLab pH 1110 Operation Manual

Ph meter
Table of Contents

Advertisement

pH value
Displaying the calibration data
24
14.
When measuring without temperature sensor: Enter the temperature of
the buffer with <><>.
15.
Start the measurement with <ENTER> .
The pH value of the buffer solution is displayed.
The measured value is checked for stability (stability control).
The [AR] status display flashes.
16.
Wait for the measurement with stability control to be completed.
17.
Set the nominal pH value of the buffer solution with <><>.
18.
Accept the calibration value with <ENTER>.
The calibration record is displayed.
5.2.5

Calibration points

Depending on the number of buffer solutions used, the meter determines the
following values and calculates the calibration line:
Values deter-
Calibration
1-point
Asy
2-point
Asy
Slo
3-point
Asy
Slo
You can display the slope in the units, mV/pH or % (see section
5.2.6 C
ALIBRATION DATA
5.2.6

Calibration data

1.
Display the calibration data in the measured value display with <CAL__>.
The value for the asymmetry (ASY) is displayed.
mined
Displayed calibration data
• Zero point = Asy
• Slope = Nernst slope
(-59.2 mV/pH at 25 °C)
• Zero point = Asy
• Slope = Slo
The calibration line goes through both cal-
ibration points.
• Zero point = Asy
• Slope = Slo
The calibration line is calculated by linear
regression.
, page 24).
TruLab pH 1110
ba76155e01
03/2014