Calibration Data - Xylem SI Analytics Lab 955 Operating Manual

Conductivity meter
Hide thumbs Also See for SI Analytics Lab 955:
Table of Contents

Advertisement

Conductivity
Displaying the calibration data
22
3.
Confirm the selection of (Use Cell) with <MENU/OK> or <CAL_>.
The cell constant that was set last is displayed.
4.
Select a cell constant in a suitable range with <CAL>.
Range 0.250 ... 2.500 cm-1
Range 0.090 ... 0.110 cm-1
Permanent, 0.010 cm
5.
Set the exact cell constant within the range with <><> as necessary.
6.
Use <MENU/OK> to finish the calibration.
The adjusted cell constant is used.
5.2.6

Calibration data

1.
Show the calibration data in the measured value display with <CAL_>.
The calibrated cell constant is displayed.
U
S
E
C L
E
L
C
A
L
4
6
0 7
Cal
-1
1/cm