Guidance and Manufacturer's Declaration - Electromagnetic Immunity
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the ME Equipment is used exceeds the applicable RF compliance level above, the ME Equipment should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating
the ME Equipment.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.
Test Specifications for Enclosure Port Immunity to RF Wireless
Communication Equipment
Test
Band
Frequency
(a)
(MHz)
(MHz)
385
380-395
450
430-470
710
745
704-787
780
810
870
800-960
930
1720
1700-
1845
1990
1970
2400-
2450
2570
5240
5100-
5500
5800
5785
Service (a)
TETRA 400
GMRS 460, FRS 460
LTE Band 13, 17
GSM 800/900, TETRA 800, iDEN 820,
CDMA 850, LTE Band 5
GSM 1800; CDMA 1900; GSM 1900;
DECT; LTE Band 1, 3. 4. 35: UMTS
Bluetooth, WLAN, 802.11 b/g/n, RFID
2450, LTE Band 7
WLAN 802.11 a/n
Modulation (b)
Maximum
Power
(W)
Pulse
modulation (b)
1.8
18 Hz
FM (c)
+/- 5 kHz
2
deviation
1 kHz sine
Pulse
modulation (b)
0.2
217 Hz
Pulse
modulation (b)
2
18 Hz
Pulse
modulation (b)
2
217 Hz
Pulse
modulation (b)
2
217 Hz
Pulse
modulation (b)
0.2
217 Hz
59
Distance
Immunity Test
(m)
Level
(V/m)
0.3
27
0.3
28
0.3
9
0.3
28
0.3
28
0.3
28
0.3
9