| Product Description
Chapter 5
|
5-4
Model 3907 & 3948
Only connect CO
sources to the CO
2
than CO
might damage the unit.
2
Set the CO
Content
2
The Watlow CO
controller's upper display shows the actual CO
2
chamber. The lower display shows the CO
To set the CO
content (0% to 20%), press the Up or Down arrow keys on the Watlow
2
PM6 Controller.
CO
sample
2
alarm
port
POWER
Watlow PM6 Control
Figure 5-4
CO
Control and Indicators
2
Sample Port - The sample port is used for checking CO
chamber by an independent test instrument (such as with a Fyrite, or similar CO
instrument).
To prevent CO
loss, the sample port must be capped when it is not in use.
2
CO
Alarm - The CO
alarm is factory set to activate when the chamber CO
2
2
deviates from configured alarm set points (see "Configuration Record" on page 13-8).
When a CO
alarm occurs, the CO
2
audible alarm sounds.
The CO
alarm high and low setpoints are established through the Watlow PM6 CO
2
controller (A.LO and A.HI). Refer to the "Configuration Record" on page 13-8.
inlet. Connecting sources of other gases
2
setpoint.
2
percentage in the environmental
2
Alarm indicator on the control panel lights and the
2
content inside the
2
test
2
content
2
2