Thermo Scientific 3907 Operating Manual page 53

Environmental chamber
Table of Contents

Advertisement

Water performance data for humidity
system
Water supply
Water specification
Input water pressure
Water consumption for humidity system
(without water recirculator system)
Measured at ambient temperature of
22°C / 72°F
typical value / max. value
Electrical Data
Power line voltage (+/-10%)
Power line frequency
Power rating with all options
Measured at ambient temperature of 32°C /82°F
Max current consumption with all options
Measured at ambient temperature of 32°C / 82°F
Power Cord / Plug
Energy consumption (without humidity control)
Measured at ambient temperature of 22°C / 72°F
typical value / max. value
Energy consumption (with humidity control)
Measured at ambient temperature of 22°C / 72°F
typical value / max. value
Earthing system (e.g. 1/N/PE)
IP protection system
Protection class
Overvoltage category according to IEC
60364-4-443
Thermo Scientific
Unit
min
25°C (77°F) /
60% RH
30°C (86°F) /
Liter per day
75% RH
40°C (104°F) /
75% RH
Unit
V
Hz
W
A
0°C (32°F)
kWh
37°C (99°F)
per
60°C (140°F)
day
5°C (41°F) / 80% RH
kWh
30°C (86°F) / 75% RH
per
60°C (140°F) / 75%
day
Specifications |
Models
3907
3948
Carboy or water system
Demineralized water
Resistance = 0.5 - 1 MΩ cm
gravity to 40 psi (2,76 bar)
ca. 4.0
ca. 5.0
ca. 11.0
ca. 10.0
ca. 1.0
ca. 1.5
Models
3907
3948
230
50/60
3450
15.0
230V:
CEE 7/7
ca. 18.5
ca. 17.0
ca. 19.0
ca. 24.5
ca. 25.5
ca. 32.5
ca. 34.5
ca. 23.0
ca. 24.0
1/N/PE AC
1/N/PE AC
IP 20
I
(PE-connected)
II
Model 3907 & 3948 | 12-3
Chapter 12
230
50/60
3450
15.0
230V:
CEE 7/7
ca. 20.5
ca. 18.0
ca. 21.0

Advertisement

Table of Contents
loading

This manual is also suitable for:

3948

Table of Contents