Thermo Scientific 3907 Operating Manual page 47

Environmental chamber
Table of Contents

Advertisement

See
Manual
Section
--
Inspect door latch, hinges and
door gasket seal.
8
Check air exchange ventilator
caps for adjustment; open or
close as required.
9
Perform a complete decontam-
ination procedure. Wipe down
interior, shelves, side panels
with disinfectant. Rinse every-
thing well with sterile water.
13
Verify and document all cali-
brations, at minimum.
10
Inspect and clean the humidity
generator, at minimum.*
5
Clean drip pan and drain lines.
9
Clean refrigeration system con-
denser.*
8
Verify defrost cycle for below
10°C operation.
--
Change filters (under normal
conditions).
--
Replace filter elements on the
dryer, if applicable. †
--
Rebuild solenoid valves on the
dryer, if applicable. †
--
Replace/repack the desiccant
towers on the dryer, if applicable.
--
Replace the check valve and
o-rings on the dryer, if applicable.
Thermo Scientific
Action
3 Months
*Qualified service personnel only ~ Regular monitoring routines of the various levels in your unit is encouraged.
† Refer to Puregas Dryer manual included in shipping box.
Yearly
2 Years
Between Experi-
ments. More fre-
quent deconta-
mination may be
required, depend-
ing on use and
environmental con-
ditions.
Depending upon
the quality of water
used in the
humidification
system, it may be
necessary to clean
more frequently.
Maintenance |
Chapter 10
5 Years
6 Years
Model 3907 & 3948 | 10-3