Instrument Base / Cross Stage; Brake Cap; Cross Stage Covers - Zeiss IOLMaster 700 Service Manual

Hide thumbs Also See for IOLMaster 700:
Table of Contents

Advertisement

IOLMaster 700

3.4 Instrument Base / Cross Stage

3.4.1 Brake Cap

Precondition:
The measurement head covers have been
removed.
See section 3.2.1, page
Procedure:
1) Using a fingernail or small slotted screwdriver
remove the small plastic cap from the
measurement head brake cap.
2) Loosen the screw underneath using a slotted
screwdriver.
3) Remove the brake cap.
Reassembly:
1) Reassemble in the reverse order.

3.4.2 Cross Stage Covers

Precondition:
The measurement head covers have been removed.
The brake cap has been removed.
Procedure:
1) Position measurement head towards
operator side. Remove the six screws
(Torx 15) on the patient side cross stage
cover and remove cover.
2) Using joystick, raise measurement head to
highest position (clockwise). Remove the
operator side cross stage cover.
Reassembly:
1) Reassemble in the reverse order.
SM-30-4060
Version 03, 2014-10-13
3.
Small Plastic Cap
Fig. 44 – Brake Cap Removal
See section 3.2.1, page
See section 3.4.1, page 21.
Fig. 45 - Cross stage covers
Confidential and Proprietary
3 - 21
Parts Removal Replacement
Screw
3.
Screws
Brake Cap
Screws
Service Manual
Order No. 2075-325

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

2075-325

Table of Contents