Adjustment and Calibration
S101 – Display device info
Description:
Retrieves and displays the following device
information:
– Serial number of IOLMaster 700
– Serial numbers of hardware components
(Measurement head, Laser, PC)
– Operating hours of measurement head and
laser
– Firmware version measurement head
– Firmware version PI
– Error codes of measurement head (if an error
is reported)
– Differences between local and head device-
settings
In addition, a Reset-command can be sent to the
measurement head (performs the start-up
function test).
S105 – Update measurement head firmware
Description:
Use this function to upload a new firmware
version onto the measurement head.
An external 'bootloader'-program, firmware file
and config file are required and must be saved in
the following location:
"ExternalPrograms\BootLoaderMcHead\".
This use-case starts the preconfigured
bootloader-program in a new window. The rest
of the work is completed automatically.
Service Manual
Order No. 2075-325
Fig. 7 – S101 Display device info
Fig. 8 – S105 Update measurement head firmware
(1)
Confidential and Proprietary
4 - 6
IOLMaster 700
SM-30-4060
Version 03, 2014-10-13