Additional programs
Program
Special 93°C-10'
Drain
Rinse
Demineralized rinse
Program selection depending on the accessories used
Upper basket
Carrier
2 injector
with spray arm
modules
for various inserts
Area of application
Program for cleaning and thermal disinfection at 199°F (93°C) with
10 minutes' temperature holding time.
The wash water is not pumped away until after disinfection is
complete.
To drain away wash water, e.g., after a program has been canceled
(see "Operation/Canceling a program").
Program for rinsing the wash cabinet, for flushing out brine (see
"Water softener/Adding reactivation salt"), or for rinsing heavily
soiled wash items, e.g., for pre-rinsing soiling, residual disinfecting
agent, to prevent items drying out, and to prevent incrustation
before running a full load. Cold water is used for rinsing, holding
time: 1 Min
Program for rinsing the wash cabinet and for rinsing wash items
with DI water, holding time: 3 Min.
Lower basket
Carrier
for various inserts
Water quantity
2 injector
modules
+ 2.0 to 2.5 L
A 303
(+ 1 module)
Program chart
Program
Normal plus
Standard
Intensive
Inorganic
Organic
Plastics
Mini
Oil program
Injector Plus
Pipettes