Frequently asked questions
Problem
Fault 578

Process-related faults and messages

Problem
Change HEPA filter
Drying during program deacti-
vated
Wrong code entered
Test program: Test object
can now be removed
Program cancelled
Program continued
140
Possible cause and solution
The peak-load cut-out has lasted longer than 3 hours.
Have your electrical system and your energy
management system tested by a suitably qualified
person.
Possible cause and solution
The maximum permitted operating hours for the HEPA filter
have been reached.
Replace the HEPA filter with a new one.
Reset the operating hours counter for the HEPA filter.
Drying cannot be selected at the start of a program
because drying is not available for the selected program.
Start the program without drying.
Or
Have the drying parameters for this program adjusted by
Miele Service.
The PIN code entered is not the same as the code saved.
Enter the PIN code again.
Report the loss of the PIN code to Miele Service.
This is not a fault.
A test program is running to check performance. At certain
points in the program, the sequence is interrupted so that
samples can be taken.
Take a sample.
or
Wait. The program will continue automatically in approx.
30 seconds.
or
Continue the program without delay by pressing the
Start/Stop button.
This is not a fault.
A program which was running was canceled by the user.
 The wash cabinet interior can be very hot.
When the door is opened, hot steam and process
chemicals can escape. Protective measures for personal
safety must be observed.
This is not a fault.
The process of canceling a program was not completed.
The program which was running continued without
interruption.