Program Settings; Adjusting Program Settings; Program Structure; Program Header - Miele PG 8593 Operating Instructions Manual

Laboratory glassware washer for laboratory glassware and utensils
Table of Contents

Advertisement

Program header

Adjusting program settings

The program settings should be adjusted to suit technical
requirements and the wash items.
Additional specialist knowledge is required to alter program settings
and this should therefore be undertaken only by experienced users or
by Miele Service.
Changing program parameters on a validated machine will require a
new performance validation.

Program structure

Each program is subdivided into program blocks which run one after
another. A program consists of at least one and a maximum of 11
program blocks. Each block can occur only once in a program.
The so-called program header is placed above the program blocks
and contains general program settings. Individual wash block
parameters are also globally activated or deactivated here.
Rinse arm monitoring
-
It is possible to monitor spray arm rotation in selected wash blocks.
Conductivity Limit
-
The conductivity of the water in the final rinse phase is monitored
using a conductivity measuring module (LFMMc).
Water volume change
-
The water intake quantity can be increased or reduced in each
program. The setting is then valid for all program blocks including
water intake.
Drainage time
-
If the on-site drainage system is insufficient to drain the waste
water from the wash cabinet within the time allocated, the drainage
time can be increased.

Program settings

99

Advertisement

Table of Contents
loading

Table of Contents