Spray Arm Monitoring/Conductivity/Wash Pressure - Miele PG 8593 Operating Instructions Manual

Laboratory glassware washer for laboratory glassware and utensils
Table of Contents

Advertisement

Frequently asked questions

Spray arm monitoring/conductivity/wash pressure

Problem
Spray arm monitoring - Upper
spray arm: Spray arm blocked
or excessive foaming
or
Spray arm monitoring - Lower
spray arm: Spray arm blocked
or excessive foaming
or
Spray arm monitoring - mobile
unit spray arm 1 - : spray arm
blocked or excessive foaming
Conductivity level too high:
Actual value: µS/cm Max
value: µS/cm
Conductivity level exceeds
limit
Conductivity module requires
calibration
Conductivity module commu-
nication error
144
Possible cause and solution
The set spin speed has not been reached.
– Wash items are obstructing the machine or basket spray
arms.
Arrange the wash items so that the spray arms can turn
easily and start the program again.
– The relevant spray arm is blocked.
Clean the spray arm.
Check whether the filters in the wash cabinet are clean
and correctly inserted.
Start the program again.
– Wash pressure is too low due to a heavy build-up of
foam.
Follow the instructions regarding foam build-up in
"Chemical processes and technology".
Rinse
Start the
program to clean the wash cabinet.
Then reprocess the wash items again.
Carry-over of conductive substances during reprocessing.
Check the process.
Empty or faulty water softener or demineralization system.
Check external water softener or demineralization
systems.
Reactivate the systems if necessary.
Work on the on-site water supply.
Contact a qualified plumber.
Water connections transposed.
Observe the markings on the water connections (see
"Connection to the water supply").
If the level drops below the measuring range, the
conductivity cannot be determined.
Contact Miele Customer Service.
The conductivity meter must be recalibrated.
Contact Miele Customer Service.
The connection to the conductivity meter has been
interrupted.
Contact Miele Customer Service.