Turn the microfine filter in the direction of the arrow and remove it
together with the coarse filter.
Press the catches towards each other and pull the coarse filter
upwards to remove it.
Remove the fine filter which sits loosely between the coarse filter
and the microfine filter.
Remove the flat filter last.
Clean the filters.
Re-insert the filter combination in the reverse order. Ensure ...
- ... that the flat filter sits flat in the base of the wash chamber.
- ... that the coarse filter has securely clicked into place in the
microfine filter.
- ... that the microfine filter is tightly screwed in as far as it will go.
If a cleaning interval was set for the filters in the wash chamber, this
interval must be reset after cleaning; see "Settings /Filter
maintenance.
Maintenance measures
123