OPMI Lumera
i on floor stand
Functional test before use
Checklist
CAUTION
CAUTION
Version 9.2
G-30-1720-en
Use the following list to check the function of the device before every opera-
tion (with no patient!).
Failure of functions possible!
If defects in the device are identified during surgery, there is a risk of injuries
to the patient.
•
Perform a functional test prior to each use of the device.
•
Do not use the device if it is defective or malfunctioning.
Failure of functions!
If the failure of a function impairs your work, do not use this device for safety
reasons!
•
Repair the cause (see section "What to do in the event of malfunctions")
or contact ZEISS Service.
•
Check zoom
The buttons on the foot control panel have been configured
(see page 112).
The zoom speed has been set on the touchscreen (see page 116).
•
Check focus
The buttons on the foot control panel have been configured
(see page 112).
The focusing speed has been set on the touchscreen (see page 116).
•
Check XY coupling
The speed has been set on the touchscreen (see page 116).
•
Check unobstructed movement of the surgical microscope
The ease of movement has been adjusted using the friction adjust-
ment knob on the stand (see page 51).
All cables and accessories have been installed in such a way that the
surgical microscope can be positioned safely without causing any
damage.
Operation