Operation
Function test before use
CAUTION
Page 166
Risk of injury!
Incorrect configuration settings may lead to unexpected behavior of the
device and may cause injury to the patient.
•
Always check the settings of the selected user profile before using the
system.
For safety reasons, the device must not be used if a malfunction is noticed
before the start of a surgical procedure.
•
Correct the malfunction (see "What to do in the event of malfunctions")
or contact ZEISS Service.
Always check the function of the system before surgery (without patient!)
using the following checklist:
Prerequisites:
Check that the correct rated voltage has been set on the suspension
system.
Check that all cables are connected and pose no trip hazards.
Check that the light guide has been connected.
There is enough room for the accessories installed.
The OPMI protective cover has been removed.
•
Turn on the system using the power switch of the suspension system.
Checking the surgical microscope
Check the zoom
•
Press the appropriate release button on the foot control panel.
Check the focus
•
Press the appropriate release button on the foot control panel.
Friction adjustment of the surgical microscope
Check that the friction of the surgical microscope's rotation has been
adjusted as required using the friction adjustment knob on the suspension
system.
Speeds of the microscope functions
Check that the speeds of the microscope functions have been adjusted as
required on the suspension system.
S88 / OPMI Lumera T
Version 8.0
G-30-1682-en