Rins Insulation Resistance Method I (With Voltage) - Fluke SMFT-1000 Calibration Manual

Multifunction pv analyzer
Hide thumbs Also See for SMFT-1000:
Table of Contents

Advertisement

SMFT-1000
Calibration Manual
R
Insulation Resistance Method I (With Voltage)
INS
To verify:
1. Set all equipment and the DUT to standby mode.
2. Connect shielded test leads from DUT red connector to the 532xA LO Port and DUT green
connector to the 532xA HI Port.
3. Connect shielded test leads between DUT red connector to the + voltage output of the
2290-5 and DUT blue connector to the - voltage output of the 2290-5.
Figure 6. R
4. Turn the rotary switch on the DUT to R
5. Push
to select R
6. Set the 532xA calibrator to output the first resistance test point and set the voltage output
of the DUT to the first voltage setting (see
16
Insulation Resistance Method I (With Voltage) Connections
INS
Fluke 5320A
I mode.
INS
LO
,
L
Power Supply
Fluke SMFT-1000
.
INS
Table
6).
,
L
+/-