Advertisement

8. Record the indication on the DUT in the test record.
9. Set the 552xA calibrator to output the next voltage test point (see
10. Repeat steps
7
11. Set the 552xA calibrator to standby mode.

AC/DC Current (SMFT only)

To verify:
1. Set all equipment and the DUT to standby mode.
2. Connect low-thermal EMF test leads (see
a. Connect DUT green connector to NORMAL red port on the 552xA and DUT yellow
connector to the 552xA black port under NORMAL red port.
b. Disconnect all other test leads.
3. Turn the rotary switch on the DUT to P
4. Push
until Current shows in highlight on the left-side menu.
5. Set the 552xA calibrator to output the first current test point (see
The 552xA output is voltage, not current, for this part of the calibration.
through
9
for all test points.
Figure 12. AC/DC Current Connections (SMFT Only)
Fluke 5522A
Figure
12):
.
AC/DC
5522A
CALIBRATOR
STBY
OPR
EARTH
EXGRD
7
8
9
4
5
6
1
2
3
/ +
0
Fluke SMFT-1000
Note
Multifunction PV Analyzer
Performance Tests
Table
11).