SMFT-1000
Calibration Manual
R
Insulation Resistance Method I (No Voltage)
INS
To verify:
1. Set all equipment and the DUT to standby mode.
2. Connect shielded test leads from DUT blue connector to the 532xA LO Port and DUT green
connector to the 532xA HI Port.
3. Connect shielded test leads between DUT red connector and DUT blue connector.
4. Disconnect all other test leads. See
Figure 5. R
5. Turn the rotary switch on the DUT to R
6. Push
button to select R
7. Set the 532xA calibrator to output the first resistance test point and set the voltage output
of the DUT to the first voltage setting (see
12
Figure
Insulation Resistance Method I (No Voltage) Connections
INS
Fluke 5320A
I mode.
INS
5.
LO
,
L
Fluke SMFT-1000
.
INS
Table
5).
,
L
+/-
METER