Fluke SMFT-1000 Calibration Manual page 13

Multifunction pv analyzer
Hide thumbs Also See for SMFT-1000:
Table of Contents

Advertisement

4. Set the 552xA calibrator to output the first voltage test point (see
Range
5 V to 99.9 V
100 V to 1000 V
5. Push OPER on the 552xA to set the operate mode, if necessary, and wait for the calibrator
to stabilize.
6. Record the indication from the DUT in the test record.
7. Set the 552xA calibrator to output the next voltage test point (see
8. Repeat steps
5
through
9. Set the 552xA calibrator to standby mode.
Table 3. V
OC
Test Points
[V]
5.0
50.0
80.0
99.2
103
500
990
7
for all test points.
Multifunction PV Analyzer
Test Points
Display Reading [V]
Lower Limit
4.8
49.6
79.4
98.5
100
496
983
Performance Tests
Table
3).
Upper Limit
5.2
50.5
80.6
99.9
106
505
997
Table
3).
9