Fluke SMFT-1000 Calibration Manual page 15

Multifunction pv analyzer
Hide thumbs Also See for SMFT-1000:
Table of Contents

Advertisement

5. Set the current supply to output the first current test point (see
Range
0 A to 20 A
Push the Output button, if necessary to verify the current supply is in operate mode.
6. Momentarily push
Wait for the DUT to complete the measurement. Record the current indication on the DUT
in the test record.
Watch the current supply during this measurement. If it does not go into constant
current mode briefly, then the voltage setting is too low. Increase the voltage of the
current supply and retry step
mode briefly. Additionally, check a range of about 0.1 V to see if the current indication
on the DUT stays constant. A constant current indication, even with a change of 0.1 V
from the current source, indicates that there is enough voltage output to allow the
current source to enter into constant current mode.
7. Set the current supply to output the next current test point (test points are in
8. Repeat steps
6
through
9. Set the current supply to standby mode.
Table 4. I
Test Points
0.2
6.0
10.0
19.5
on the DUT.
6
until the current supply goes into constant current
7
for all test points.
Test Points
SC
[A]
Lower Limit
0.0
5.7
9.7
19.1
Note
Multifunction PV Analyzer
Performance Tests
Table
4).
Display Reading [A]
Upper Limit
0.4
6.3
10.3
19.9
Table
4).
11