Fluke SMFT-1000 Calibration Manual page 42

Multifunction pv analyzer
Hide thumbs Also See for SMFT-1000:
Table of Contents

Advertisement

SMFT-1000
Calibration Manual
Fluke SMFT-1000
3. Turn the rotary switch on the DUT to P
a. Push
until Current shows in highlight on the left-side menu.
b. Use the dial on the current clamp to zero the clamp I
The absolute value of the DC current reading on the DUT display should not exceed
0.1 A dc.
c. When the clamp is zeroed, push
d. Momentarily push
4. Set the 552xA calibrator to output the first DC power test point (see
38
Figure 15. AC/DC Power Connections
Fluke 5522A
AC/DC
to set the DUT to enter Power measurement mode.
on the DUT to enter Continuous DC measurement mode.
5522A
CALIBRATOR
PREV
STBY
OPR
EARTH
EXGRD
SCOPE
MENU
p
dBm
sec
7
8
9
m
V
Hz
n
W
F
4
5
6
k
A
C
p
1
2
3
M
F
/ +
0
ENTER
SHIFT
:
reading on the DUT display.
DC
EDIT
SETUP
RESET
FIELD
NEW
CE
REF
MEAS
MORE
TC
MODES
MULT
DIV
x
POWER
Table
14).