Detailed Specifications

For all detailed specifications:
Accuracy is given as ±([% of reading] + [number of least significant digits]) at 18 ºC to 28 ºC, with relative humidity up to
90 %, for a period of one year after calibration.
In the 4 ½-digit mode, multiply the number of least significant digits (counts) by 10. AC conversions are ac-coupled and
valid from 3 % to 100 % of range. The Meter is true rms responding. AC crest factor can be up to 3 at full scale, 6 at half
scale. For non-sinusoidal wave forms add -(2 % Rdg + 2 % full scale) typical, for a crest factor up to 3.

AC Voltage Function

Function
Range
K
600.0 mV
[2,4]
6.000 V
60.00 V
600.0 V
1000 V
Low pass filter
[1]
Below 10 % of range, add 12 counts.
[2]
The Meter is a true rms responding meter. When the input leads are shorted together in the ac functions, the Meter may display a
residual reading between 1 and 30 counts. A 30 count residual reading will cause only a 2-digit change for readings over 3 % of range.
Using REL to offset this reading may produce a much larger constant error in later measurements.
[3]
Frequency range: 1 kHz to 2.5 kHz.
[4]
A residual reading of up to 13 digits with leads shorted, will not affect stated accuracy above 3 % of range.
[5]
Specification increases from -1% at 200 Hz to -6% at 440 Hz when filter is in use.

DC Voltage, Resistance, and Conductance Function

Function
L
F
mV
e
nS
When using the REL ∆ function to compensate for offsets.
[1]
[2]
Add 0.5 % of reading when measuring above 30 MΩ in the 50 MΩ range, and 20 counts below 33 nS in the 60 nS range.

Temperature

-200 ºC to +1090 ºC
-328 ºF to +1994 ºF
[1]
Does not include error of the thermocouple probe.
Accuracy specification assumes ambient temperature stable to ± 1 ºC. For ambient temperature changes of ± 5 ºC, rated accuracy
[2]
applies after 1 hour.
6
Resolution
45 - 65 Hz
0.1 mV
± (0.7 % + 4)
0.001 V
0.01 V
± (0.7 % + 2)
0.1 V
1 V
± (0.7 % + 2) ± (1.0 % + 4)
Range
6.000 V
60.00 V
600.0 V
1000 V
600.0 mV
600.0 Ω
6.000 kΩ
60.00 kΩ
600.0 kΩ
6.000 MΩ
50.00 MΩ
60.00 nS
Temperature
30 - 200 Hz 200 - 440 Hz 440 Hz - 1 kHz
± (1.0 % + 4)
+1 % + 4
[5]
-6 % - 4
Resolution
0.001 V
0.01 V
0.1 V
1 V
0.1 mV
0.1 Ω
0.001 kΩ
0.01 kΩ
0.1 kΩ
0.001 MΩ
0.01 MΩ
0.01 nS
Resolution
0.1 ºC
1 % + 10
0.1 ºF
1 % + 18
Accuracy
1 - 5 kHz
± (2.0 % + 4)
± (2.0 % + 4)
unspecified
unspecified
unspecified
Accuracy
± (0.05 % + 1)
± (0.05 % + 1)
± (0.05 % + 1)
± (0.05 % + 1)
± (0.1 % + 1)
[1]
± (0.2 % + 2)
± (0.2 % + 1)
± (0.2 % + 1)
± (0.6 % + 1)
± (0.6 % + 1)
[2]
± (1.0 % + 3)
[1,2]
± (1.0 % + 10)
[1,2]
Accuracy
[1]
5 - 20 kHz
± (2.0 % + 20)
[3]
unspecified
unspecified
unspecified