11. Momentarily push
Wait for the DUT to finish both the + and - measurements. Record the + and - indications on
the DUT in the test record. Also record the nominal value displayed on the 532xA as it may
be slightly different from the intended nominal. If applicable, recalculate the limits based on
the nominal value and the published specifications.
Ignore the red X or green
12. Set the 532xA calibrator to output the next resistance test point (see
13. Repeat steps
10
14. Set the 532xA calibrator to standby mode.
V
(Open-Circuit Voltage)
OC
To verify:
1. Set all equipment and the DUT to standby mode.
2. Connect low-thermal EMF test leads from DUT red connector to 552xA NORMAL port and
DUT blue connector to the 552xA black port below NORMAL port. Disconnect all other test
leads (see
Figure
3. Turn the rotary switch on the DUT to V
8
on the DUT.
. They do not indicate a pass or fail for this calibration.
through
12
for all test points.
3).
Figure 3. V
(Open-Circuit Voltage) Connections
OC
Fluke 5522A
Note
5522A
CALIBRATOR
STBY
OPR
EARTH
EXGRD
7
8
9
4
5
6
1
2
3
/ +
0
Fluke SMFT-1000
/I
.
OC
SC
Table
2).
PREV
SCOPE
MENU
p
dBm
sec
EDIT
m
V
Hz
SETUP
RESET
FIELD
n
W
F
NEW
CE
k
A
C
REF
p
MEAS
MORE
M
F
TC
MODES
MULT
DIV
ENTER
SHIFT
x
POWER

Advertisement

Table of Contents
loading

Table of Contents