2.2 – Compressed air supply
The M20-EPS requires compressed air supply to work. The connect must the made
using the back quick-fit connection (6mm OD hose), as shown below. Inlet air pressure
must be between 40 and 110 psi (2.7 to 10.2 bars).
Picture 4 – Compressed air
inlet port connection
WARNING:
 Apply clean and low humidity compressed air. Contamination by water
and/or oil will cause equipment malfunction. A 5µm particulate filter,
followed by a 0.01µm coalescent filter is recommended to remove any
residual contamination that may be present in the compressed air line;
 Air filters are recommended;
NEVER use air lubrication system in the air supply;
 Make sure all air connection is secure and leak-free.
 For safety reasons, an emergency button for blocking the
compressed air supply must be available to the operator. Review the
safety guidelines described in this manual.
12
Picture 5 – Measurement needle
connection port (DO NOT connect
to the compressed air network)