Section IV
Calibration of DUT's
To
minimize the calculation error, the inherent dissipation of the capacitor should
be
0.001
or
below.
When
using
polystyrene
or silvered
mica
type
capacitors
(dissipation
factor is generally very low), the residual factors will not affect the
derivation of accurate dissipation factors. If dissipation of capacitor alone is greater
than 0.001, the effective value of the DUT is calculated in accord with the following
equation:
Ds =Dc + Dr
(Dr Dc, Dr € 0.01)
where,
Ds is actual dissipation factor of DUT.
Dc is calculated D value (excludes inherent dissipation).
Dr is inherent dissipation of capacitor.
Compensate
the
dissipation
factor
for the
measuring
frequencies
of individual
4262A
being tested or calibrated.
Convert the D value of the calibration frequency
to that of the actual
4262A
measuring
frequency
in accord with the following
equations:
f
Dm: D value at 4262A
-m-
% =
measuring frequency.
fm
Ds:
D value at calibration
frequency.
Dm =X-Ds
fm: 4262A measuring
_ fm
frequency
—Fw
TR
fc:
Calibration frequency.
Note
To accurately measure frequencies fm and fc,
use a reciprocal counter or calculate reciprocal
number of period.
[CALIBRATION EQUIPMENT]
The
recommended
model
and
required
performance
of calibration equipment is
listed below:
Instrument
Required Performance
Recommended Model
Capacitance Accuracy: 0.1%
Capacitance Bridge | Dissipation Factor Accuracy:
GR 1620-A
0.1% (Resolution 0.0001)
.
.
HP 3490A
DMM
Resistance Accuracy: 0.02%
HP 3455A
Reciprocal counter
HP 5300A/5307A
Freq. Counter
| Resolution: 0.01Hz
HP 5323A
4-5