Table 3-4.
Unusual Operating Indications (Sheet 3 of 4).
Indication:
Cause
of trouble:
Cause
A.
Effect of Low terminal capacitance with
respect to ground.
Excessive
measurement
error.
Sometimes
the measurement
can not be
Measurement
Allowable Stray
performed when a relatively large capacitance
Frequency
Capacitance Magnitude
between Lpor terminal and ground exists,
Allowable
magnitudes
for stray capacitance
120Hz
100nF
without additional error are given in figure at
10kHz
200pF
Cause B .
.
.
.
Effect of High terminal capacitance with
UNKNOWN
respect to ground.
The stray capacitance
cvarp
o
.
will reduce test signal level applied to the
Cur
POT
HpOT
Hcur
oc s -
oC 1S+
sample
measured
during capacitance
measurement.
This decrease
in signal level
will not produce
an additional error
even
when
measurement
signal level is reduced to
N
a third of its nominal level.
It is neccessary,
of course,
that special care be taken to use
Re
AN
the proper test signal level when a device is
H
d whose
parameters may
be affected
L
Sample
measured
w h o s e
p
y
¢
T
p
:'L.
by the test signal level.
Display fluctuations
{7
{7
may
sometimes
appear.
Indication:
What to do:
Internal resistance of a battery can not be
1.
Connect
sample battery
(observe polarity)
measured.
.
as illustrated.
2.
Batteries up to 40V are measured under
no load conditions,
UNKNOWN
3.
If battery voltage exceeds 4V,
set
GUARD
LcunDcmAsL_Pov
HeoT
DCBIAS'HCUR
DC
BIAS
to
EXT
4,
Since the internal resistance of a battery
is relatively
low,use the four- terminal
measurement configuration.
Battery
-
3-25