Equipment Required; Testrecord; Calibration Cycle; Accuracy Checks - HP 4262A Operating And Service Manual

Digital lcr meter
Hide thumbs Also See for 4262A:
Table of Contents

Advertisement

Model 4262A
Section IV
Paragraphs 4-1 to 4-8
SECTION
1V
PERFORMANCE
TESTS
4-1. INTRODUCTION.
4-2. This
section provides
the check
procedures
to
verify
the
4262A
specifications
listed
in
Table
1-1. All tests can be performed without ac-
cess to the interior of the instrument. A simpler
operational test is presented in Section III under
Self
Test
(paragraph
3-5).
The
performance
test procedures in this section can also be used to
do
an incoming
inspection
of the instrument and
to verify whether the instrument meets its speci-
fied performance after troubleshooting or making
adjustments. If specifications
are found to be out
of limits, check that controls are properly set, and
then proceed to adjustments or troubleshooting.
Note
Allow
a
15-minute
warm-up
and
stabilization period before conduct-
ing any performance test.
4-3. EQUIPMENT REQUIRED.
4-4.
Equipment
required for the performance tests
is listed in Table 1-4 Recommended
Test Equip-
ment
in
Section
I.
Any
equipment
whose
characteristics equal the critical specifications given
in the table may
be substituted
for the recom-
mended model(s).
Accuracy
checks
in
this
section
use
standard
LCR
components as the samples to be connected
to the 4262A.
Accessories
16361A
and 16362A
can
be utilized for this purpose. These accessory
models
are DUT
(device under test) boxes from
which the desired component can be selected and
connected to the 4262A through cables by use of a
rotary
switch.
If
models
16361A/16362A
are
unavailable,
use
the
discrete
components
rec-
ommended
in Table 4-1.
Note
All components
used
as standards
should
be calibrated by an instru-
ment
whose
specifications
are
traceable to NBS, PTB, LNE, NRC,
JEMIC,
or
equivalent
standards
group; or all components should be
calibrated directly by an authorized
calibration
organization
such
as
NBS.
The
calibration
cycle should
be determined by the stability spec-
ification for each component.
45. TEST RECORD.
4-6. Results of the performance tests may be tabu-
lated on the Test Record at the end of these proce-
dures. The Test Record lists all the tested specifica-
tions
and
their
acceptable
limits. Test results re-
corded
at incoming
inspection
can
be used for
comparison
in periodic maintenance and trouble-
shooting and after repairs or adjustments.
4-7. CALIBRATION CYCLE.
4-8. This instrument requires periodic verification
of performance. Depending on the use and envir-
onmental
conditions,
the
instrument
should
be
checked
with
the following performance
tests at
least once every year. To maximize the "up time"
of the
instrument,
the recommended
preventive
maintenance
frequency for the 4262A
is twice a
year.
4-1