Model 4262A
Table
3-61. Remote Message Coding.
D
D
|
|
CLASS
0
o
8
76
543
21
DCL
device
clear
ucC
X
0010100
GET
group
execute
trigger
AC
X
00
01000
GTL
go
to
local
AC
X
00
00O0O01
LLO
local
lock
out
ucC
X000
10001
MLA
my
listen
address
AD
X0
1
LLLLYL
5
4
3
2
1
MTA
my
talk
address
AD
X1
0TTTTT
5
4
3
2
1
OTA
other
talk
address
AD
( OTA=TAGNMTA)
SDC
selected
device
clear
AC
X0
000100
SPD
serial
poll
disable
ucC
X
00110
001
SPE
serial
poll
enable
ucC
X0011000
STB
status
byte
ST
S
XS
SSSS
S
UNL
unlisten
AD
X01
11111
UNT
untalk
AD
X1
011111
CLASS
UC
! Universal
Command
AC
! Addressed
Command
AD
! Address
ST . Status
Byte
Section
III
Table 3-61
3-35