PERFORMANCE TESTS
13)
Again
enables
acceptance
of SRQ
from
device
because SRQ is disabled when Status Byte signal
transfer is completed (re to Para 3-70).
14)
After service subroutine is completed, return to
the step that follows
step
7, 8, or 9 as appro-
priate to main programming sequence.
Serial Poll
1
L Step 10
*STB = Status Byte
Stores *STB
in register B
Step 15
READ DATA
Step 11
Checks SRQ
YES/NO line
of *STB
Step 12
Print
Setp 13
Interrupt Enable
Return
Step 14
Figure 4-11 SRQ Service Routing.
[RESULTS]
Calculator
prints
LCR
and
DQ values of the sample measured
by 4262A
(test fre-
quency 10kHz). Verifies that 4262A SRQ lamp lights momentarily. Press calculator
RUN
button
again
to repeat checks. If calculator prints OTHER
DEVICE
SRQ,
interface is faulty.