Zero Detector & Apao - HP 4262A Operating And Service Manual

Digital lcr meter
Hide thumbs Also See for 4262A:
Table of Contents

Advertisement

Model 4262A
Section V
Figure 5-11
ADJUSTMENT
(2) ZERO DETECTOR
& APAO
ADJUSTMENT.
T
==l
=
.
—em—
.o
©OBH
98
"@@
'%
'e @BOO
BBO
7
7
DUT
BOX
PROCEDURE A.
(a)
Figure 5-11. Zero Detector & APAO Adjustments.
Note
If DUT
Box
is available, use procedure A. If
not, use procedure B.
(b)
a.
Adjust
"ZOF"
potentiometer A14R1
for 1000 counts £1 count on 4262A
LCR
display.
b.
Adjust
"APAO"
potentiometer
A14R15
for .000 to .001
count on 4262A
DQ
display.
c.
Set 4262A TEST SIGNAL control successively to each test frequency and test signal
level shown
in Table
5-5 and confirm
that DC voltmeter readings are within 0 to
+4
volts
at each control
setting. Also confirm
that 4262A
LCR
display
and
DQ
display are within the tolerances described in steps a and b.
Table 5-5. TEST SIGNAL Settings.
Frequency
Low Level
120Hz
off
1kHz
off
10kHz
off
120Hz
on
1kHz
on
10kHz
on
Note
If result
of confirmation
check is not satis-
factory,
readjust
"OFS-4"
potentiometer
Al3R67
for
any
voltage
between +1 volt and
+3 volts
to satisfy the
requirements of step
c.
If this adjustment fails to bring the volt-
age
at A14TP3
to within
its tolerance orto
satisfy the confirmation check, refer to Sec-
tion VIII for troubleshooting,
5-15