Unusual Operating Indications - HP 4262A Operating And Service Manual

Digital lcr meter
Hide thumbs Also See for 4262A:
Table of Contents

Advertisement

Model 4262A
Section
III
Table
3-4
Table 3-4.
Unusual Operating Indications (Sheet 1of 4).
Indication:
Cause
of trouble:
A, Same
sample
sometimes
shows quite
different values between
PRL
and
SER CIRCUIT MODE
measurements,
B. The decimal point moves
and measure-
ment
unit
changes.
A and/or B may occur in the following cases:
Resistance
of low loss inductor
or
capacitor being measured in
R FUNCTION.
Indication:
What to do:
A. Do not set CIRCUIT MODE
to AUTO.
Set CIRCUIT MODE to a PRL
or SER
setting that shows a valid display.
B. Set LCR RANGE
to MANUAL.
Manually
settle the instrument
on an appropriate range.
Cause
of trouble:
The displayed value fluctuates on minimum
capacitance,
maximum
inductance
or
maximum
resistance
ranges
in either PRL
or SER
circuit modes.
Here
are some
of
the reasons why this
happens:
A .
A large
size sample
is being measured,
B .
A high voltage power line or similar
exists near the 4262A, .
UNKNOWN
GUARD
LroT
HroT
H CuR
0C Blas =
DC BIAS +
D
D
i
Cx
-.-.",_-..
Metal
Case
Lcunr
What to do:
1.
Enclose
sample
in metal case.
Connect
case electrically to 4262A
GUARD
terminal
as illustrated.
2.
Use shielded cable for connection
between
sample and the instrument.
Connect
cable
shield to GUARD.,
3-23