Model 4262A
8-53.
PRODUCT
SAFETY
CHECKS.
8-54.
WARNING
WHENEVER
IT
APPEARS
LIKELY
THAT
SAFETY
PROTECTIVE
PRO-
VISIONS
HAVE
BEEN
IMPAIRED,
THE
APPARATUS
SHALL
BE
MADE
INOPERATIVE
AND
BE
SECURED
AGAINST
ANY
UNINTENDED
OPER-
ATION,
THE
PROTECTION
IS
LIKELY
TO BE
COMPROMISED
IF,
FOR EXAMPLE:
-- THE
APPARATUS
SHOWS
VISI-
BLE DAMAGE.
-- THE
INSTRUMENT
FAILS
TO
PERFORM
THE INTENDED MEAS-
UREMENT.
-- THE UNIT HAS UNDERGONE
PRO-
LONGED
STORAGE
UNDER
UN-
FAVORABLE
CONDITIONS.
-- THEINSTRUMENT HAS SUFFERED
SEVERE TRANSPORT
STRESS.
The
following
five
checks
are
recommended
to verify the product
safety of the 4262A LCR Meter
(these checks
may
also
be
done to check for product
safety after troubleshooting and repair).
When such
checks
are needed,
perform the following:
1,
Visually inspect
interior
of instrument for any
signs
of abnormal, internally
generated
heat,
such as discolored printed
circuit boards or
components,
damaged
insulation,
or
evidence
of arcing.
Determine
and remedy
cause of any
such condition.
Using
a suitable
ohmmeter,
check
resistance
from
instrument
enclosure
to
ground
pin
on
power
cord plug.
The
reading
must
be less
than
0.5
ohm.
Flex
the
power
cord
while
making this measurement
to determine
whether
intermittent discontinuities exist.
Check
GUARD
terminal
on
front
panel
using
procedure
(2).
Disconnect
instrument
from
power
sSource.
Turn
power
switch to
on.
Check resistance
from
instrument
enciosure to line and
neutral
(tied together).
The minimum acceptable re-
sistance
is two
megohms.
Replace
anycom-
ponent which fails or causes a failure.
Check
line
fuse
to
verify
that
a correctly
rated fuse is installed.
Section VIII
Paragraphs 8-53 and 8-54
8-29