Table 5-2. Factory Selected Components.
Reference
:
Designator
Nominal Value Range
Effect on Performance
HP P/N: 0757-0440, R:FXD 7.5kQ
Changes test signal level. If signal level
Al11R16
» HP P/N: 0698-3259, R:FXD 7.87kQ
is too high, use less resistance; if too
HP P/N: 0757-0441, R:FXD 8.25kQ
low, use more resistance.
)
.
Minimizes dissipation measurement
A12C1
gg g/fi; 8%68'8158' C;IEXD 6800??
error on *100nF (100uF) and *10uH
(Para. 5-23.)
' HP P/
: 0120'0
60, C:F§g 8200p
(10mH) ranges at 10kHz measurement.
IN:
-0161, C:
10CCOpF
Refer to Paragraph 5-23 (2).
Minimizes dissipation measurement
Al12C2
» HP P/N: 0140-0190, C:FXD 39pF
error on 100pF (100nF) and *10mH
(Para. 5-23)
HP P/N: 0160-2201, C:.FXD 51pF
(10H) ranges at 10kHz measurement.
Refer to Paragraph 5-23 (4).
Changes adjustment range for dissi-
pation measurement error on *10pF
(10nF) and 100mH ranges at 10kHz
measurement. Refer to Paragraph 5-23
(3).
A12C3
» HP P/N: 0121-0059, C:VAR 2 - 8pF
(Para. 5-23)
HP P/N: 0121-0036, C:VAR 5.5 - 18pF
HP
P/N:
0160-2199,
C:FXD
30pF
Rejects parasiti(; oscillatiqn of
Al2C14
» HP P/N- 0160-2307:
C:FXD 47pF
Al2U2 OP AMP in measuring
10m§Q
resistor at 10kHz.
Changes adjustment range for dissi-
Al3C1
» HP P/N: 0121-0059, C:VAR
2 - 8pF
pation measurement error on all ranges
(Para. 5-23)
HP P/N: 0121-0036, C:VAR 5.5 - 18pF
at 10kHz measurement. Refer to para-
graph 5-23 (1).
A13C5
» HP P/N: 0160-2251
5. 6pF
Changes the phase delay of
HP P/N: 0160-2253
6.8pF
Al13U3B OP AMP.
A13C23
» HP P/N: 0160-
Changes the feedback signal
/ N :
0160-0134 220pF
amount of AL3U5B OP AMP.,
> gg g;fi gigggggg' 85'))%3 éggg
Eliminates switching transient noise
Al4C5
HP P/N: 0160-2202, C:FXD 75pF
If\§°m.A114 g'lhas? detecl'i" O"EP"t-
HP P/N: 0160-2203, C:FXD 91pF
ominal value 1s usually used.
Note: éomponent marked ( p ) in table is usually used.
*
Ranges in PRL mode for capacitance and in SER mode for inductance.
Values in ()
are ranges in SER mode for capacitance and in PRL mode
for inductance.