3.3 ELECTRICAL POWER CONNECTION
Do not use any extension lead.
Make sure that electric power is constant. Make sure that electric power
is constant. Please note that the compressor requires a start-up current
between 25 and 35 A. The electric circuit at the installation site must be
inspected by an electrical engineer to ensure that it meets the
requirements for a smooth operation of the instrument.
• This should be examined during installation of the unit by a
• Use a dedicated fuse for the unit
• Do not connect another unit to the same power circuit.
• Before turning on the instrument, check if the voltage of the mains
3.4 WORKING WITH KNIVES AND DISPOSABLE BLADES
Microtome knives and disposable blades have extremely sharp edges and this can lead
to injuries.
Please be extremely careful when handling microtome knives and disposable blades.
Do not place microtome knives or disposable blades at unsecured areas.
Never position microtome knives or disposable blades with the sharp edge
pointing upwards.
Store blades in a covered container. Use a container that has guides to
hold the blades rigid.
Never try to catch a falling microtome knife.
Always insert the specimen first and then the microtome knife or
disposable blade.
To avoid compression or knife marks, ensure that the blade is clean.
7
competent person
supply is identical with the name plate of the unit
Manual_TN60_2021-09-001.00_EN_Tanner.docx