TANNER Scientific Cryostat TN60 Operating Instructions Manual page 17

Table of Contents

Advertisement

7.4 AUTOMATIC DEFROST, DISINFECTION AND POWER SAVING SETTINGS
Press
to open the Settings menu.
Select Program settings and press
menu for setting automatic defrost and power saving
times.
Enter PIN Code 5792.
PIN Code is usually restricted to service technicians or
instrument administrators
Select defrosting programs 1 to 3 and set a time for
the automatic defrosting programme.
(e.g. each day at 12:00 h and at 23:00 h in this
example)
Select temperature increase start and set a starting
time for the automatic power saving mode.
(e.g. start at 21:00 h in this example)
Select temperature increase end and set a ending
time for the automatic power saving mode.
(e.g. end at 04:00 h in this example)
17
>
to open the
>
>
>
>
Manual_TN60_2021-09-001.00_EN_Tanner.docx
Chamber temp
Quickfreeze
UV disinfection
Light
TIME
TUE
13 05 05
Language
Display version
Program settings
Enter PIN
0000
Defr. Start 1
Defr. Start 2
Defr. Start 3
Disinf. Req.
Disinf. delay
Incr. start
Incr. end
_ _ _ _ _ 5
Incr. days
Disinf. delay
Incr. start
Incr. end
_ _ _ _ _ 5 6
Incr. days
-20 °C
Off
Auto
English
5 µm
12:00
23:00
__:__
__:__
10 h
21:00
04:00
6
10 h
21:00
04:00