7.5 List of available software settings
Chamber temperature
Object cooling
Object temperature
Quick-freeze
UV disinfection
Light
Light duration
19
-35 °C to ambient
The recommended working temperature is -15 °C to -25 °C
ON / OFF
If the instrument is equipped with the optional
object cooling system, its function can be activated
or deactivated.
-50 °C to 0 °C
If the instrument is equipped with the optional
object cooling system, its function can be activated
or deactivated.
ON / OFF
If the instrument is equipped with the optional fast
cooling peltier system, its function can be activated
or deactivated.
The default setting for automatic deactivation of the
quick freeze is set to 30 minutes. The timer for
automatic deactivation can be set by an authorized
service technician.
ON /OFF
If the instrument is equipped with the optional UV-
C disinfection system, its function can be activated
or deactivated.
The default setting for automatic deactivation is set
to 30 minutes. The timer can be set by an
authorized service technician.
ON / OFF / AUTO
The light can be set on or off. If AUTO has been
selected, the light will switch off after the time
being set under Light duration.
0 - 99 minutes