10.3 RECOMMENDED MAINTENANCE AND SERVICE SCHEDULE
Remove frozen section waste from the cryostat chamber after every day
Daily
of usage.
Disinfection if necessary.
Weekly
Cleaning of blade holder (see section 10.3).
Cleaning of window.
Check of temperature sensor in cryochamber.
Monthly
Cleaning of cryostat ventilation (see section 10.4).
Yearly
Complete Service
• check of all functions
• complete defrost
• removing of microtome
• oiling of movable parts
• check of cooling system
• check of driving system
• complete disinfection / cleaning / drying
Manual_TN60_
2021-09-001.00
(performed by authorized Tanner Scientific service technician)