Select temperature increase days and set the days at
that the automatic power saving mode will be active
for 24 hours.
(e.g. Fridays and Saturdays in this example)
1-Monday 2-Tuesday 3-Wednesday 4-Thursday 5-Friday 6-Saturday 7-Sunday
Settings for disinfection parameters:
The disinfection cycle works in a semi-automated
way.
First select a time when the system will ask for a
disinfection cycle.
(e.g. at 12:00 h the instrument will ask for a
disinfection cycle).
Select the days in which a disinfection cycle can be
activated.
(e.g. Monday and Tuesday in this example)
1-Monday 2-Tuesday 3-Wednesday 4-Thursday 5-Friday 6-Saturday 7-Sunday
Select the delay time at which the disinfection cycle
will be activated.
(e.g. 10 hours)
After confirmation the disinfection cycle will start at
10 p.m.
Manual_TN60_
2021-09-001.00
_EN_Tanner.docx
Disinf. delay
Incr. start
Incr. end
>
Incr. days
>
Disinfec. Req.
Disin. Days
Disin. Delay
Disinfec. Req.
>
Disin. Days
Disin. Delay
Disinfec. Req.
Disin. Days
>
Disin. Delay
18
1 2 _ _ _ _ _
21:00
04:00
_ _ _ _ 5 6 _
12:00
1 2 _ _ _ _
10 h
12:00
1 2 _ _ _ _
10 h
12:00
1 2 _ _ _ _
10 h