The device is delivered with a
defrosting water collecting container,
which must be installed before the
device is put into operation.
First mount the metal bracket for the
defrosting water container on the right
side of the device. Loosen the screws
and hang the bracket in the screws.
Now tighten the screws.
Place the defrosting water container in
the metal holder and connect the
defrosting water pipe and the level
sensor to the device.
It is necessary to install the laterally defrost container, otherwise there
i
may be icing on the evaporator! We recommend filling the container
with 500 ml disinfectant before initial operation.
After installation, wait at least 4 hours before switching on the instrument. Failure to
comply may result in severe damage to the instrument.