Operation

5.11.2 SENSOR-SETUP

WIPE
VERIFY
CALIBRATION
STANDARD
OFFSET
CAL. FACTOR
CAL. CONFIG
SET CAL DEFLT
CONFIGURATION
LOCATION
MEAS UNIT
CLEANING
RESPONSE TIME
LOGGERINTERV.
RESOLUTION
SET DEFAULTS
44
MAIN MENU
DIAG
DIAG
TEST/MAINT
Selection as per
CAL.CONFIG OUTPUT
MODE
Selection as per
CAL.CONFIG OUTPUT
MODE
FAC. STANDARD
OFFSET
OUTPUT MODE
CAL INTERVAL
mg/l, FNU, NTU, TE/F,
EBC
Confirmation prompt
SELECT SENSOR
Verification using CVM
module
Calibration with standard
solution
Zero point setting
HOLD
ACTIVE
TRANSFER
SELECTION
SET DEFAULTS Instrument
number
SET DEFAULTS
FNU
SET DEFAULTS
12 h
SET DEFAULTS
15 s
SET DEFAULTS
10 min
SET DEFAULTS
0,xxx
客服:400 688 9809
Initiates a wiping action
Verification with dry standard
Procedure during calibration, menu
based
Zero point setting, menu based
0.50 to 2.00
-0.100 to +0.100 TRBFNU
Behaviour of the outputs during
calibration or zero point setting
Adjustable from 0 to 365 days
Reset to the default calibration
up to 16-character name possible
10 min, 20 min, 30 min,
2 h, 6 h, 12 h, 10:00
0 to 60 s
1-30 min
0,xxx or 0,xxxx (<1 FNU)
Reset to default configuration for all
menu commands listed above.