Section 6
6.1 Maintenance schedule
6.2 Cleaning measuring chamber
50
Maintenance
It has been possible to reduce the scope of the maintenance to be performed by
the user to only a few actions. Clearly summarised in a table and described in
detail in the following sections, these actions can be quickly and straightforwardly
performed by qualified personnel.
Maintenance task
Clean measuring chamber:
ULTRATURB sc
ULTRATURB plus sc
Replace wiper profile
(only plus version)
Replace desiccant
Monitor test equipment (CVM dry
standard)
Check zero point
Check gradient
The cleanliness of the measuring chamber is crucial for correct measured results.
This cylindrical optical measuring chamber has a rotating wiper with three wiper
profiles (only on plus version); these prevent normal soiling of the optical system
as it starts to build up. With very stubborn deposits, manual cleaning with a
suitable cleaning agent (e.g. citric acid) is recommended.
Note: Observe safety regulations and wear safety clothing!
•
Safety glasses
•
Gloves
•
Overall
Select
1.
2.
3.
4.
5.
6.
Interval
Dependent on the substances contained
in the water
after 1200 cycles
Every 2 years
Every 2 years
(factory test with certification)
Dependent on the substances contained
in the water
At least once a year
Menu level
MAIN MENU
SENSOR-SETUP
SELECT SENSOR
(if there is more than one sensor)
TEST/MAINT
MAINT.PROC.
CLEAN MEAS.CHAMBER
客服:400 688 9809
Accept