Section 5 Operation; Sensor Setup; Sensor Data Logging - Hach 6120118 User Manual

Digital pc sc and rc sc 3/4 -inch combination ph/orp sensor
Table of Contents

Advertisement

Section 5

5.1 Sensor Setup

5.2 Sensor Data Logging

5.3 Sensor Diagnostics Menu for pH and ORP
SELECT SENSOR (if more than one sensor is attached)
STATUS
ERROR LIST
WARNING LIST
5.4 pH Sensor Setup Menu
SELECT SENSOR (if more than one sensor is attached)
CALIBRATE (Main Menu Item)
1 POINT AUTO
2 POINT AUTO
1 POINT MANUAL
2 POINT MANUAL
TEMP ADJUST
DEFAULT SETUP
CONFIGURE
EDIT NAME
SELECT MEASURE
DISPLAY FORMAT
TEMP UNITS
LOG SETUP
Operation
When a sensor is initially installed, the serial number of the Digital Gateway will be
displayed as the sensor name. To change the sensor name refer to the following
instructions:
1. Select Main Menu.
2. From the Main Menu, select SENSOR SETUP and confirm.
3. Highlight the appropriate sensor if more than one sensor is attached and confirm.
4. Select CONFIGURE and confirm.
5. Select EDIT NAME and edit the name. Confirm or cancel to return to the Sensor
Setup menu.
The sc1000 controller provides a data log for each sensor.
The sc100 provides three data logs (one for each sensor and one for calculated values).
The data logs store the measurement data at selected intervals. The event log stores a
variety of events that occur on the devices such as configuration changes, alarms, and
warning conditions. The data logs are stored in a packed binary format and the event logs
are stored in a CSV format. The logs can be downloaded through the digital network port,
service port, or the IrDA port. DataCom is needed for downloading logs to a computer.
See
section 7.1 on page
See
section 7.2 on page
Calibration with a single buffer — normally pH 7.
Calibration with two buffers — normally pH 7 and pH 4 or 10.
Calibration against a single known sample.
Calibration against two samples, both with a known pH.
Adjust the displayed temperature by up to ± 15 °C.
Restores the system to the original factory calibration.
Enter up to a 10-digit name in any combination of symbols and alpha or numeric characters.
Select the appropriate measurement units to display.
Select the measurement resolution (xx.xx pH or xx.x pH).
Choose from the displayed options (°C or °F).
Choose SENSOR INTERVAL to set the sensor log interval or select TEMP INTERVAL to set the
temperature log interval.
35.
35.
25