Instrument Specifications; External Power Source Requirements - In-situ Aqua TROLL 700 Operator's Manual

Hide thumbs Also See for Aqua TROLL 700:
Table of Contents

Advertisement

Operating temperature
Storage temperature
Dimensions
Weight
Wetted materials (sonde and
sensors)
Environmental rating
Max pressure rating
Communication
Reading rate
LCD screen
External power voltage
External power current
(Dependent on display/wiping)
Cable
Hex screw driver
Software Interface
Certifications
Warranty
Notes

External Power Source Requirements

Current from an external power source must not exceed 4 amps.
1-970-498-1500

Instrument Specifications

-5 to 50° C (23 to 122° F)
Nitrate and Ammonium Sensors: 0 to 40° C (32 to 104° F)
Chloride Sensors: 0 to 50° C (32 to 122° F)
Components without fluid: -40 to 65° C (-40 to 149° F)
pH/ORP sensors: -5 to 65° C (23 to 149° F)
Nitrate and Ammonium Sensors: 0 to 40° C (32 to 104° F)
Chloride Sensors: 0 to 50° C (32 to 122° F)
Diameter: 7.21 cm (2.84 in) O.D. without bumpers;
8.23 cm (3.24 in) O.D. with bumpers
Length: 48.67 cm (19.16 in) (includes connector)
With bail: 61.67 cm (24.28 in)
2.25 kg / 4.96 lbs. (includes all sensors and bail)
Buna-N, Noryl
, Nylon, Polyphenylsulfone, Polycarbonate, Acetal,
TM
EPDM/Polypropylene TPV, FKM Fluoroelastomer, Titanium, Flourocarbon
Coating, Ceramic, Acrylic Adhesive Film
IP68 with all sensors and cable attached. IP67 with sensors removed or
cable detached
Up to 350 PSI; Ammonium/Nitrate up to 30 PSI
RS485/MODBUS, SDI-12, Bluetooth®
1 reading every 2 seconds for one parameter, no wipe
Integrated display shows status of sonde, sensor ports, power voltage
and connectivity.
8-36 VDC (required for normal operation)
Sleep: < 0.2 mA typical
Measurement: 40 mA typical, 75 mA Max
Vented or non-vented polyurethane or vented Tefzel®
0.050 in. (1.3 mm)
Android/iOS: VuSitu Mobile App (see app store for OS requirements)
Data Services: HydroVu
CE, FCC, WEEE, RoHS, UKCA Compliant
2 year - sonde, sensors (excluding ISE sensors)
1 year - Nitrate, Chloride, and Ammonium Sensors
Other - see warranty policy at www.in-situ.com
Specifications are subject to change without notice. Android is a
trademark of Google, Inc. Bluetooth is a trademark of Bluetooth SIG, Inc.
Noryl is a registered trademark of SHPP Global Technologies B.V.
9
www.in-situ.com

Advertisement

Table of Contents
loading

This manual is also suitable for:

Aqua troll 800