Storing the Sensor
Prior to installing the cap, store the cap above freezing in the container in which it was shipped. Store the
instrument in a clean, dry area between -5 and 60°C.
After installing the cap, store the instrument in a clean, dry area above freezing.
Maintenance
Clean and inspect your instrument regularly for accurate readings and best performance.
With installing the instrument for the first time, inspect the instrument weekly to determine the best
cleaning frequency for your deployment site.
If the sensor measurements begin to drift from expected values, clean the sensor and calibrate. If calibration
cannot be performed, or you are receiving invalid measurements or errors, replace the RDO cap.
How to Clean the Instrument
Unscrew and remove the
guard.
Disposal
Recycle the instrument according to local regulations. Dispose of calibration solutions as described in the
Safety Data Sheet accompanied with the calibration solution.
In the UK and EU, Partech can provide recycling of your old Partech equipment and can sometimes recycle
other manufacturers' equipment when replaced with Partech equipment. Trade-in options may be available
for new equipment. Contact Partech for more information.
1-970-498-1500
Storage and Maintenance
Gently clean the cap with
a soft cloth. Use Alconox
or vinegar to remove
build-up.
Replace the guard and
calibrate the instrument.
7
Do not use organic
solvents to clean cap.
Never wet the lens area
under the cap. Use only
the supplied lens cloth to
clean lens.