Operating temperature
Storage temperature
Dimensions
Weight
Wetted materials (sonde and
sensors)
Environmental rating
Max pressure rating
Communication
Reading rate
LCD screen
External power voltage
External power current
Cable
Hex screw driver
Software interface
Certifications
Warranty

External Power Source Requirements

Current from an external power source must not exceed 4 amps.
1-800-833-7958

Instrument Specifications

-5° to 50° C (23° to 122° F)
Nitrate and Ammonium Sensors: 0 to 40° C (32 to 104° F)
Chloride Sensors: 0 to 50° C (32 to 122° F)
Components without fluid: -40° to 65° C (-40° to149°F)
pH/ORP sensors: -5 to 65° C (23 to 149° F)
Nitrate and Ammonium Sensors: 0 to 40° C (32 to 104° F)
Chloride Sensors: 0 to 50° C (32 to 122° F)
Length: 46 cm (18.145") (includes connector)
Diameter: 4.7 cm (1.860")
With bail: 59cm (23.25")
0.978kg / 2.15 lbs. (includes instrument, sensors, restrictor and
bumpers)
Polyphenylsulfone, Polycarbonate, Acetal, EPDM/Polypropylene TPV,
FKM Fluoroelastomer, Titanium, Flourocarbon Coating, Ceramic,
Inconel, Acrylic Adhesive Film, Nylon, Polyurethane Adhesive,
Graphite, PC/PMMA Blend, Acrylic, Sapphire, PVC, Platinum, Glass
IP68 with all sensors and cable attached. IP67 with sensors removed
or cable detached
Up to 150 PSI
Ammonium/Nitrate up to 30PSI
RS485/MODBUS, Wireless TROLL Com, Bluetooth®
1 reading every 2 seconds
Integrated display shows status of sonde, sensor ports, power voltage
and connectivity. BlueTooth may be disabled through the hidden
menu
8-36 VDC (required for normal operation)
Sleep: < 0.2 mA typical
1
Measurement: 40 mA typical, 75 mA Max
Vented or non-vented polyurethane or vented Tefzel®
0.050 in. (1.3 mm)
Android/iOS: VuSitu Mobile App (see app store for OS requirements)
Windows: Win-Situ 5
Data Services: HydroVu
CE, FCC, WEEE, RoHS, UKCA Compliant
2 year - sonde, sensors (excluding ISE sensors)
1 year - Nitrate, Chloride, and Ammonium Sensors
Other - see warranty policy at www.in-situ.com