Thermo Scientific RC 12BP Plus Instruction Manual page 58

Laboratory centrifuge
Table of Contents

Advertisement

7
Troubleshooting
Failure message
Failure message
Problem with centrifuge
INTERNAL PROGRAM
The centrifuge cannot be operated.
ERROR
The run does not start or the centrifuge runs
down without being braked.
FREQUENCY
The centrifuge cannot be operated.
CONVERTER VOLTAGE
The run does not start or the centrifuge runs
TO HIGH
down without being braked.
7-6
RC 12BP+
Possible causes and cures
The centrifuge detected a program fault.
Restart the centrifuge.
If an error message appears again, turn of the centrifuge.
The emergency lid release enables you to retrieve your
samples.
Pull out the power supply plug.
Inform Customer Service.
The line voltage exceeds the tolerance or failure of the braking
resistance.
Restart the centrifuge.
If an error message appears again, turn of the centrifuge.
The emergency lid release enables you to retrieve your
samples.
Pull out the power supply plug.
Inform Customer Service.